• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - SPENCER RX CYTOLOGY BRUSH; ENDOSCOPIC CYTOLOGY BRUSH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - SPENCER RX CYTOLOGY BRUSH; ENDOSCOPIC CYTOLOGY BRUSH Back to Search Results
Model Number M00545000
Device Problem Bent (1059)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/24/2014
Event Type  malfunction  
Event Description
It was reported to boston scientific corporation that an rx cytology brush was opened for use in the bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2014.According to the complainant, during preparation a functional test was performed and difficulty was encountered during extension and retraction of the brush.It was reported that the bristled portion of the device was bent, and the brush fibers were bending as well, causing the brush to appear as though it was oriented differently.The procedure was completed with another rx cytology brush.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be good.
 
Manufacturer Narrative
Reported event of brush bent.Although the suspect device has been received, the evaluation has not been completed.Therefore, the cause of the reported malfunction has not been determined.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
 
Manufacturer Narrative
Investigation results: visual evaluation found that the pull wire was bent approximately 3.2cm from the distal end; the bristled portion of the brush was not bent.Some blue paint was peeled off at the distal tip of the catheter, and the handle was bent approximately 4.0cm from the thumb ring.Functional evaluation found that the device was difficult to actuate.The brush could only partially retract due to the bend in the pull wire.It is most likely that the pull wire was bent during handling of the device, which would impact its ability to extend and retract due to increased resistance.A review of the device history record (dhr) confirmed that the device met all material, assembly, and product specifications at the time of release to distribution.A search of the complaint database confirmed that no other complaints exist for the specified batch.
 
Event Description
It was reported to boston scientific corporation that an rx cytology brush was opened for use in the bile duct during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2014.According to the complainant, during preparation a functional test was performed and difficulty was encountered during extension and retraction of the brush.It was reported that the bristled portion of the device was bent, and the brush fibers were bending as well, causing the brush to appear as though it was oriented differently.The procedure was completed with another rx cytology brush.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be good.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RX CYTOLOGY BRUSH
Type of Device
ENDOSCOPIC CYTOLOGY BRUSH
Manufacturer (Section D)
BOSTON SCIENTIFIC - SPENCER
780 brookside drive
spencer IN 47460
Manufacturer (Section G)
BOSTON SCIENTIFIC - SPENCER
780 brookside drive
spencer IN 47460
Manufacturer Contact
ingrid matte
100 boston scientific way
marlborough, MA 01752
5086834000
MDR Report Key3688779
MDR Text Key4301231
Report Number3005099803-2014-01431
Device Sequence Number1
Product Code FDX
Combination Product (y/n)N
PMA/PMN Number
K930348
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/24/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/19/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/12/2015
Device Model NumberM00545000
Device Catalogue Number4500
Device Lot Number16376814
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/27/2014
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/17/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/17/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-