Catalog Number 5100004000 |
Device Problem
Device Displays Incorrect Message (2591)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 02/24/2014 |
Event Type
malfunction
|
Event Description
|
The tps handpiece cord was returned to stryker instruments for service.During functional testing by a service technician, it was found that the device caused a bias current message to be displayed on the console, signaling a condition occurred in which the device has the potential to run without user activation.No medical intervention and no adverse consequences were reported with this event.As this event occurred during testing at the manufacturer facility, there was no patient involvement and no delay to a surgical procedure.
|
|
Manufacturer Narrative
|
It was confirmed the cable caused a bias current error to be displayed by the service technician through functional evaluation.During the device inspection, a quality engineer determined the cable¿s internal wiring was damaged.The attachment is not a repairable device and will not be returned to the user facility.
|
|
Manufacturer Narrative
|
Failure analysis is in progress; additional information will be submitted once the quality investigation is completed.
|
|
Event Description
|
The tps handpiece cord was returned to stryker instruments for service.During functional testing by a service technician, it was found that the device caused a bias current message to be displayed on the console, signaling a condition occurred in which the device has the potential to run without user activation.No medical intervention and no adverse consequences were reported with this event.As this event occurred during testing at the manufacturer facility, there was no patient involvement and no delay to a surgical procedure.
|
|
Search Alerts/Recalls
|