• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMERICAN MEDICAL SYSTEMS (MN) AMS RETROARC RETROPUBIC SLING SYSTEM; MESH, SURGICAL, POLYMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMERICAN MEDICAL SYSTEMS (MN) AMS RETROARC RETROPUBIC SLING SYSTEM; MESH, SURGICAL, POLYMERIC Back to Search Results
Device Problem Folded (2630)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/08/2014
Event Type  Injury  
Event Description
Additional information from the ams representative on (b)(6) 2014 indicated she was present at the event.It was also reported that during the retroarc mesh implant the mesh was positioned under the urethra.The sheaths were noted to be sticking on the midline area.The physician applied pressure to the sheaths to unstick the midline area and in the process the mesh was put under a large amount of tension causing a curling effect on the edge of the mesh.The physician made a choice to remove the sling and replace with another manufacturer's device.
 
Event Description
It was reported that during a retroarc sling implant the "mesh rolled under urethra during implantation." the device had to be removed and another manufacturer's device was implanted to complete the procedure.There were no patient complications reported in association with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS RETROARC RETROPUBIC SLING SYSTEM
Type of Device
MESH, SURGICAL, POLYMERIC
Manufacturer (Section D)
AMERICAN MEDICAL SYSTEMS (MN)
10700 bren road w
minnetonka MN 55343
Manufacturer (Section G)
AMERICAN MEDICAL SYSTEMS (IRELAND)
athlone business & tech park
garrycastle, dublin road
co. westmeath
EI  
Manufacturer Contact
sharon zurn
10700 bren road w
minnetonka, MN 55343
9529306000
MDR Report Key3691290
MDR Text Key4247402
Report Number2183959-2014-00086
Device Sequence Number1
Product Code OTN
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/08/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/20/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/08/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-