Catalog Number 3500000710
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/25/2014
Event Type  malfunction  
Event Description
It was reported via repair work order that the recliner could not support weight due to the backrest coming off its brackets.No patient was affected and no adverse consequence or clinically relevant delay in treatment was reported.
 
Event Description
It was reported via repair work order that the recliner could not support weight due to the backrest coming off its brackets and one of the attachment knobs being in the wrong location and the other being loose.No patient was affected and no adverse consequence or clinically relevant delay in treatment was reported.
 
Manufacturer Narrative
It was initially reported that the recliner could not support weight due to the backrest coming off its brackets.Further investigation also determined that one of the attachment knobs was in the wrong location and the other was loose.It was further determined that the customer assembled the unit upon delivery and did not follow the backrest installation and removal process as detailed in manual.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TREATMENT RECLINER
Type of Device
CHAIR, EXAMINATION AND TREATMENT
Manufacturer (Section D)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer (Section G)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer Contact
pravin betala
3800 east centre avenue
portage, MI 49002
2693292100
MDR Report Key3693732
MDR Text Key4293859
Report Number0001831750-2014-02550
Device Sequence Number1
Product Code FRK
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/25/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/21/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number3500000710
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received02/25/2014
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-