Catalog Number 3500000710 |
Device Problem
Loose or Intermittent Connection (1371)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 02/25/2014 |
Event Type
malfunction
|
Event Description
|
It was reported via repair work order that the recliner could not support weight due to the backrest coming off its brackets.No patient was affected and no adverse consequence or clinically relevant delay in treatment was reported.
|
|
Event Description
|
It was reported via repair work order that the recliner could not support weight due to the backrest coming off its brackets and one of the attachment knobs being in the wrong location and the other being loose.No patient was affected and no adverse consequence or clinically relevant delay in treatment was reported.
|
|
Manufacturer Narrative
|
It was initially reported that the recliner could not support weight due to the backrest coming off its brackets.Further investigation also determined that one of the attachment knobs was in the wrong location and the other was loose.It was further determined that the customer assembled the unit upon delivery and did not follow the backrest installation and removal process as detailed in manual.
|
|