AUDIO SERVICE GMBH AUDIO SERVICE BLUETOOTH TRANSMITTER; HEARING AID REMOTE CONTROL TRANSMITTER Back to Search Results
Catalog Number 10420243
Device Problem Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problem Electric Shock (2554)
Event Date 01/29/2014
Event Type  malfunction  
Event Description
Hearing aid dispenser in (b)(6) informed audio service by phone that a patient asserted they received an electrical shock by touching the remote control bluetooth transmitter.The patient consulted a doctor because of problems with their heartbeat.Dispenser indicated in a follow-up phone call that the patient was absolutely fine and had no impairment of health.
 
Manufacturer Narrative
Investigation of the transmitter determined: visual inspection of all plugs, cables, sockets and housings of the transmitter, audio cable and power supply were undamaged.Output voltage of the usb plug measured 5.165v/dc.No higher voltage measured.Transmitter was switched on and the blue led was blinking correctly.Transmitter was connected to the power supply and the red led indicated correct charging process.The transmitter was paired via bluetooth with a rcu.The blue led blinked and after a few seconds the pairing was done correctly.An audio signal was delivered via an audio cable to the bluetooth transmitter.The audio signal was audible in a connected hearing aid.No evidence was found for causing an electric shock.It was not possible to duplicate the complaint.Further evaluation of the specifications indicates the transmitter does not have the potential to send out electrical energy to a patient when touching the transmitter.The internal voltage is a maximum 5v when connected to a charger, and 4.2v when a battery is used.An electrostatic shock is possible only if the device is connected to another device (e.G., tv set) where a connection to a power supply net exists.If an electrostatically charged person touches the transmitter, the energy from the person may be shortcutted via the tv cable to the power supply net.Although the transmitter is not affected, the person may get a non-hazardous electrostatic shock the same way as they would when getting close to the water tap or other conducting appliance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUDIO SERVICE BLUETOOTH TRANSMITTER
Type of Device
HEARING AID REMOTE CONTROL TRANSMITTER
Manufacturer (Section D)
AUDIO SERVICE GMBH
zeppelinstr. 9
herford 3205 1
GM  32051
Manufacturer Contact
zeppelinstr. 9
herford 32051
221380742
MDR Report Key3701594
MDR Text Key4332509
Report Number3000246090-2014-00001
Device Sequence Number1
Product Code PGQ
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Consumer
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/21/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/22/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number10420243
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer02/03/2014
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/03/2014
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-