Device 1 of 2.Reference mfr report # 1627487-2014-01190.It was reported the patient is without stimulation and experienced jolting sensations in her back.Reportedly, impedances are low.X-ray were taken and results are normal.Follow-up revealed the physician may currently consider other therapies with the patient.However, the future course of action may be to revise the system.Further investigation identified 4 contacts with low impedance and a possible lead pullout.Reference mfr report # 1627487-2014-26105 regarding the ipg.
|