It was reported that after a da vinci surgical procedure, the wire of the fenestrated bipolar forceps instrument was observed to stick out of the jaw of the instrument.Nothing reportedly fell into the patient.The planned surgical procedure was completed and no patient harm, adverse outcome or injury was reported.
|
The instrument was returned and evaluated.Per the customer reported complaint, failure analysis investigation found that the pitch up cable was frayed at the distal clevis hub.The frayed strands stuck out at the wrist.The other cables at the wrist were undamaged.No other damage was found.The customer reported complaint does not itself constitute a mdr reportable event; however, the frayed cable could likely cause or contribute to an adverse event, if the malfunction were to recur.
|