Device Problem
High impedance (1291)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 03/06/2014 |
Event Type
malfunction
|
Event Description
|
It was reported that device diagnostics resulted in high impedance (dc dc code - 7).It was reported that the device was programmed off after observing the high impedance.It was reported that there was no trauma that is believed to have caused or contributed to the high impedance.X-rays were sent to manufacturer for review.No discontinuities with the vns system were seen in the x-rays.It was unable to determine whether or not the lead pin was fully inserted into the generator header due to the orientation of the generator in the images.Attempts to obtain additional relevant information have been unsuccessful to date.No surgical intervention has been performed to date.
|
|
Event Description
|
An implant card was received indicating that the vns patent underwent generator and lead replacement surgery on (b)(6) 2014 due to lead discontinuity.The explanting facility discarded the explanted devices; therefore, no analysis can be performed.Review of the available programming and diagnostic history showed normal diagnostic results through (b)(6) 2011.
|
|
Manufacturer Narrative
|
Device failure is suspected, but did not cause or contribute to a death or serious injury.
|
|
Manufacturer Narrative
|
Device manufacturing records and available programming and diagnostic history were reviewed.Review of manufacturing records confirmed that the lead passed all functional tests prior to distribution.
|
|