CYBERONICS, INC. LEAD MODEL 302 Back to Search Results
Device Problem High impedance (1291)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/06/2014
Event Type  malfunction  
Event Description
It was reported that device diagnostics resulted in high impedance (dc dc code - 7).It was reported that the device was programmed off after observing the high impedance.It was reported that there was no trauma that is believed to have caused or contributed to the high impedance.X-rays were sent to manufacturer for review.No discontinuities with the vns system were seen in the x-rays.It was unable to determine whether or not the lead pin was fully inserted into the generator header due to the orientation of the generator in the images.Attempts to obtain additional relevant information have been unsuccessful to date.No surgical intervention has been performed to date.
 
Event Description
An implant card was received indicating that the vns patent underwent generator and lead replacement surgery on (b)(6) 2014 due to lead discontinuity.The explanting facility discarded the explanted devices; therefore, no analysis can be performed.Review of the available programming and diagnostic history showed normal diagnostic results through (b)(6) 2011.
 
Manufacturer Narrative
Device failure is suspected, but did not cause or contribute to a death or serious injury.
 
Manufacturer Narrative
Device manufacturing records and available programming and diagnostic history were reviewed.Review of manufacturing records confirmed that the lead passed all functional tests prior to distribution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEAD MODEL 302
Type of Device
LEAD
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058 770
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key3724441
MDR Text Key4328024
Report Number1644487-2014-00880
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/06/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/03/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Date Manufacturer Received05/22/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age12 YR
-
-