LIFE TECHNOLOGIES HOLDINGS PTE. LTD. APPLIED BIOSYSTEMS 7500 FAST X REAL-TIME PCR Back to Search Results
Catalog Number 4407205
Device Problem Overheating of Device (1437)
Patient Problem No Information (3190)
Event Date 02/07/2014
Event Type  malfunction  
Event Description
A customer reported an overheated plastic smell coming from the instrument, applied biosystems 7500 fast dx (cat.No.# 4407205, serial no.(b)(4)).No pt involvement reported.
 
Manufacturer Narrative
Device is an instrument and is not implanted/ explanted.The instrument was investigated on site.The complained issue could not be reproduced.While checking the instrument components, a slight discoloration on the lamp cable to the lamp holder connector was observed.It was repaired on site by replacing the lamp holder, lamp cable and lamp.Oq/pv was performed after the repair and the instrument passed required specifications.Device intended use: the applied biosystems 7500 fast dx real-time pcr instrument with the sds software version 1.4 is a real-time nucleic acid amplification and detection system that measures nucleic acid signals from reverse transcribed rna and converts them to comparative quantitative readouts using fluorescent detection of dual-labeled hydrolysis probes.The 7500 fast dx real-time pcr instrument is to be used only by technologists trained in laboratory techniques, procedures, and on use of the analyzer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
APPLIED BIOSYSTEMS 7500 FAST X REAL-TIME PCR
Manufacturer (Section D)
LIFE TECHNOLOGIES HOLDINGS PTE. LTD.
singapore
SN 
Manufacturer (Section G)
LIFE TECHNOLOGIES HOLDINGS PTE. LTD.
Manufacturer Contact
kelli tanzella
3175 staley rd.
grand island, NY 14072
7167743122
MDR Report Key3727142
MDR Text Key4325395
Report Number3003673482-2014-00002
Device Sequence Number1
Product Code NSU
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K082562
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Consumer,Company Representative
Reporter Occupation Not Applicable
Type of Report Initial
Report Date 02/07/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/27/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number4407205
Device Lot Number275030204
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received02/07/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/01/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown