|
Model Number M-5491-01 |
Device Problems
Air Leak (1008); Device Alarm System (1012); Failure to Sense (1559)
|
Patient Problem
Air Embolism (1697)
|
Event Date 03/12/2014 |
Event Type
malfunction
|
Event Description
|
During a procedure, it was reported that air bubble was noted in the tubings.The air was in the vascular system.After the procedure, the patient was checked and found stable, there were no consequences.On march 19th, received additional information requested from bwi representative stating that the pump sensor did not detect the air bubbles; therefore, no alarm error was displayed.Based on the additional information, the bubble sensor failed to detect the air bubbles.This could lead a potential risk to the patient.Therefore, this complaint became reportable.(b)(4).
|
|
Manufacturer Narrative
|
The equipment investigation is still in progress.(b)(4).
|
|
Manufacturer Narrative
|
(b)(4).During a procedure, it was reported that air bubble was noted in the tubings.The air was in the vascular system.After the procedure, the patient was check and found stable, there were no consequences.On march 19th, received additional information requested from bwi representative stating that the pump sensor did not detect the air bubbles therefore, no alarm error was displayed.The equipment was evaluated and found physical damage as well as the bu sensors found defective.The bubble sensors, door window, ground strap and pcb interface were replaced.The equipment was also subjected to pm, safety and functional testing and all tests passed.The system is ready for use.The device history records was reviewed and no anomalies were found related to this complaint.In addition, the dhr review verifies that the device was manufactured in accordance with documented specification and procedures.The customer complaint was confirmed.
|
|
Search Alerts/Recalls
|
|
|