• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. GYNECARE TVT EXACT CONTINENCE SYSTEM; MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. GYNECARE TVT EXACT CONTINENCE SYSTEM; MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC Back to Search Results
Catalog Number TVTRL
Device Problem Reaction (1514)
Patient Problems Allergic reaction (1701); Itching Sensation (1943); Treatment with medication(s) (2571)
Event Type  Injury  
Manufacturer Narrative
(b)(4).A final report was received from the physician confirming the patient does not have any delayed hypersensitivity reactivity to polypropylene mesh.In addition, a review of the batch manufacturing records was conducted and the batch met all finished goods release criteria.
 
Event Description
It was reported that the patient underwent an unknown procedure in 2011 and mesh was implanted.The patient began experiencing itching and rashes immediately after placement of the device.The patient has been given medications to help treat the symptoms.The patient has undergone patch testing which was negative and has been given medications to help treat the symptoms.
 
Manufacturer Narrative
(b)(4).No conclusion can be drawn at this time.Should additional information be obtained, a supplemental 3500a form will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GYNECARE TVT EXACT CONTINENCE SYSTEM
Type of Device
MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC
Manufacturer (Section D)
ETHICON INC.
p.o. box 151, route 22 west
somerville NJ 08876 015
Manufacturer (Section G)
ETHICON SARL-NEUCHATEL
puits-godet 20 2000 neuchatel
neuchatel
SZ  
Manufacturer Contact
guillermo villa
route 22 west po box 151
somerville, NJ 08876
9082180707
MDR Report Key3731192
MDR Text Key11784746
Report Number2210968-2014-04369
Device Sequence Number1
Product Code OTN
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K132054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/14/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/07/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date07/31/2011
Device Catalogue NumberTVTRL
Device Lot Number3447680
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received04/09/2014
Was Device Evaluated by Manufacturer? No
Date Device Manufactured07/20/2010
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-