Catalog Number UNK_INS
Device Problem Break (1069)
Patient Problem No Patient Involvement (2645)
Event Date 03/10/2014
Event Type  malfunction  
Event Description
It was reported that the trinkle attachment broke off inside of the single trigger rotary handpiece during set up.There was no patient involvement, and there were no user injuries or adverse consequences.
 
Manufacturer Narrative
Upon disassembly of the handpiece, the gear train and rotary front end assembly were discovered to be worn/damaged in addition to the spindle being broken, which can be caused or contributed to by an attachment being broken off inside the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN_INSTRUMENTS_PRODUCT
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, DC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
Manufacturer (Section D)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer (Section G)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer Contact
casey metzger
4100 east milham avenue
kalamazoo, MI 49001
2693237700
MDR Report Key3731807
MDR Text Key4307408
Report Number0001811755-2014-01210
Device Sequence Number1
Product Code KIJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 03/11/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/07/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNK_INS
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/17/2014
Date Manufacturer Received03/11/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/26/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
SNGL TRGR ROTARYPN (B)(4)
-
-