Model Number 5086MRI45
Device Problem Kinked (1339)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/08/2014
Event Type  malfunction  
Manufacturer Narrative
Product event summary: the full lead was returned, analyzed and the inner insulation of the lead was extrinsically breached due to pulling/stretching/overstress.The proximal conductor of the lead became extrinsically distorted due to kinking/ buckling and the visual summary analysis of the lead indicated damage at implant.
 
Event Description
It was reported that the physician had difficulty trying to advance the lead into a venous sheath and that they felt the insulation "crimping" about 10 centimeters from the tip of the lead.The lead body was pulled near the area of the issue and the insulation straightened out.The lead was not implanted.No patient complications were reported as a result of this event.
 
Manufacturer Narrative
The information submitted reflects all relevant data received.If additional relevant information is received, a supplemental report will be submitted.(b)(4).
 
Manufacturer Narrative
Corrected information: no eval explain code.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPSUREFIX MRI SURESCAN
Type of Device
DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM DISEASE MGMT
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
nashoane fulwood-kelley
8200 coral sea st ne
mounds view, MN 55112
7635260583
MDR Report Key3743459
MDR Text Key12062259
Report Number2649622-2014-03403
Device Sequence Number1
Product Code NVN
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P090013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative,company representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 02/08/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/11/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date09/06/2015
Device Model Number5086MRI45
Device Catalogue Number5086MRI45
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/18/2014
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/22/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/19/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age00068 YR
-
-