(b)(4).The customer generated three discrepant results with the cobas taqscreen mpx test relative to serology testing.Three donor samples tested in primary pools of 6 (pp6) with the mpx test generated mpx non-reactive results.However, the three donors were (b)(6).Further confirmatory testing for (b)(6) using a recombinant immunoblot assay (riba) confirmed the three donor samples were (b)(6).Due to the discrepant results obtained, the three donor samples were not released for donations.The datafiles containing the complaint data could not be provided for review, as the results had been purged from the customer's cobas s 201 system.This mdr addresses sample #3 there was insufficient sample volume left to perform resolution testing with the cobas taqscreen mpx test to investigate.Therefore, the customer returned sample was tested with the cobas® taqman hcv test, v2.0 for use with the high pure system and a 'target not detected' (tnd) result was obtained for the sample.The lod of the cobas taqman® hcv test, v2.0 for use with the high pure system is 20 iu/ml while the lod of the (b)(6) target used by the customer (mpx test using pp6) is 66 iu/ml.This tnd result does not necessarily mean the (b)(6) target is not present in the sample but the absence of detection could also be due to the (b)(6) titer in the sample being below the lod of the (b)(6) test used.As 'target not detected' results were obtained for the sample during testing with the cobas® taqman® hcv test, genetic sequencing was performed with the remaining volume to determine if there were any sequence mismatches in the sample that may explain the result discrepancies reported in this case.The following conclusions were made from the genetic sequencing results: despite several attempts to obtain (b)(6) sequences from the customer returned sample, no viral sequences were obtained.The lack of hcv sequence does not necessarily mean that no (b)(6) virus was present in the sample.The lack of sequence information could be due to sequence heterogeneity under the primers used for sequencing or due to low viral titers.Retain testing of the complaint lot 122992 generated valid results for all controls and qc standards tested with no discrepant results obtained.The retain testing results confirm appropriate test performance and functionality of the complaint kit lot.According to the mpx package insert (m/n 05888212001, doc rev.9.0), the discrepant results obtained in this global case may be explained by: 1)mutations present in the mpx detection target region which can result in failure to detect the hcv target and/or 2) low hcv viral titers present in the sample tested.The mpx test is most sensitive during individual donor testing than during pp6 testing.This is because the limit of detection of the mpx test during individual donor testing for the hcv target is 11 iu/ml and 66 iu/ml during pp6 testing.The tnd result generated with the cobas taqman® hcv test, v2.0 for use with the high pure system indicates that it is below the lod of the test, which is 20 iu/ml.Given that mutations within the highly conserved regions of the viral genomes covered by the cobas® taqscreen mpx test primers and/or probes are rare, the most likely cause for the discrepant results is the presence of a low hcv titer in the sample during pp6 testing.Review of cdc interpretation of results of test for (b)(6) infection and further actions indicate that a test outcome of serology (b)(6) not detected is interpreted as no current (b)(6) infection as this scenario can occur in patients that have had exposure and/or treatment for (b)(6).(b)(4).
|