It was reported that during a laparoscopic partial gastrectomy the tip of the device burned, then the device stopped functioning.Another device was used to complete the procedure.There was a two minute delay, and no patient injury.This report is being filed for the findings upon device investigation.
|
Final device investigation found that the device was returned with the tissue pad melted and split, and the tip of the pad broken off and not returned with the device.Upon evaluation, the device passed all electrical testing, including a self-test and button tests.The device history record was reviewed, and no discrepancies were noted.
|