On (b)(6) 2014, it was reported by the biomed at the hospital, that the belmont rapid infuser, fms2000, exhibited an "over temperature" error message during a liver transplant.The system was replaced and the case was continued.
|
The device was returned for eval.Upon receipt, the fan guards were very dirty and there was a lot of blood on the housing, suggesting the unit was not cleaned after each use as stated in the operator's manual.Otherwise, the unit performed according to our specifications.The system shut down due to over temperature, as it is designed to do: whenever fluid from the heat exchanger reaches our specified limit, the system stops pumping and heating, closes off the line to the pt and instructs the user to restart the system with a new disposable.The volume of tubing between the temperature sensor and the pt is 45 ml so that none of the over temperature fluid can reach the pt.The disposable was discarded by the user and therefore not returned for eval.
|