Catalog Number PC0830RXC
Device Problem Defective Component (2292)
Patient Problem Vasoconstriction (2126)
Event Date 03/25/2014
Event Type  Injury  
Manufacturer Narrative
As reported by the (b)(6) registry, during a carotid artery stenting (cas) procedure a carotid artery spasm was noted during deployment of a 9x40mm precise pro rx stent.It was also noted that the stent became compressed in the patient.A second precise pro rx was deployed overlapping the first stent.Cas was performed on a 90% occluded lesion in the ostium of the left internal carotid artery of 20mm in length in a 9.0mm vessel diameter with moderate vessel tortuosity.The lesion was eccentric and moderately calcified.A 6mm extra support embolic protection device was deployed past the lesion and a 9x40mm precise stent was deployed.The stent became compressed after a carotid artery spasm occurred as depicted by an angiogram showing a narrowing at the proximal end of the stent.A second precise pro rx was deployed overlapping the first stent with a residual diameter stenosis measured at 10%.The patient was neurologically intact upon leaving the angio suite.The patient was discharged on the following day with no reported patient injury.A review of the manufacturing documentation associated with this lot presented no issues during the manufacturing process that can be related to the reported complaint.Vessel spasm is a known potential adverse event associated with any interventional procedure where devices are introduced into the vasculature and is listed in the ifu (instructions for use) as such.Local vasospasm can be caused by the outward radial force and axial friction of the stent, or by the device manipulations inherent in any procedure causing endothelial irritation.A carotid vessel spasm is a brief temporary tightening of the muscles in the vessel wall.This can narrow and briefly decrease or even prevent blood flow to the brain.This may lead to tia like symptoms such as muscular or neurological changes.Treatment of arterial spasms may include medications such as nitrates and calcium channel blockers.It was also noted that the stent became compressed after the carotid artery spasm occurred as depicted by an angiogram showing a narrowing at the proximal end of the stent.A second precise pro rx was deployed overlapping the first stent.There is no evidence that manufacturing issues contributed to the event.Review of the information suggests that patient, vessel and procedural factors may have contributed to the reported events.The product was not returned for analysis.Based on the lack of information and the inability to assign or determine a root cause no corrective actions will be taken at this time.This is one of two products involved with the reported event and are associated manufacturer report numbers 9616099-2014-00271 and 1016427-2014-00047.
 
Event Description
As reported by the (b)(4) registry, during a carotid artery stenting procedure a carotid spasm was noted during deployment of a 9x40mm precise pro rx stent which also became compressed in the patient.A second precise pro rx was deployed overlapping the previous stent.Cas was performed on a 90% occluded lesion in the ostium of the left internal carotid artery of 20mm in length in a 9.0mm vessel diameter with moderate vessel tortuosity.The lesion was eccentric and moderately calcified.A 6mm extra support embolic protection device was past the lesion and a 9x40mm precise stent was deployed and became compressed after a carotid spasm (crinkling of the artery) because the angiogram showed narrowing at the proximal end of the stent.An 8x30mm precise pro rx stent was deployed proximal to the target lesion and overlapping the previous stent.The residual diameter stenosis measured 10%.There was no presence of air bubbles noted during the procedure.The patient was neurologically intact upon leaving the angio suite.The patient was discharged on the following day without any further events.
 
Manufacturer Narrative
This device is not available for testing and evaluation.Additional information is pending and will be submitted within 30 days upon receipt.J wire, ima catheter, 5f sheath, 6f 90 cm sheath, 6mm angioguard, 8x30mm precise stent and mynx closure device.Concomitant medications: heparin was given during the procedure.Pre and post-procedure medications included aspirin and clopidogrel.This is one of two products involved with the reported event and are associated manufacturer report numbers 9616099-2014-00271 and 1016427-2014-00047.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRECISE PRO RX CAROTID STENT SYSTEM
Type of Device
SELF EXPANDING STENTS (NIM)
Manufacturer (Section D)
CORDIS DE MEXICO
circuito interior norte #1820
juarez, chihuahua 3258 0
MX  32580
Manufacturer Contact
cecil navajas
miami lakes, FL 33014
63138802
MDR Report Key3753255
MDR Text Key11755884
Report Number9616099-2014-00271
Device Sequence Number1
Product Code NIM
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030047
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Study,Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/28/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/16/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2015
Device Catalogue NumberPC0830RXC
Device Lot Number15936091
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received04/22/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/01/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age63 YR
Patient Weight63
-
-