Method: the complaint mr290v vented autofeed humdification chamber was returned to fisher & paykel healthcare in (b)(4) and was visually inspected.Results: visual inspection revealed that both of the chamber ports were deformed.The hospital reported that the mr290v chamber was set up with an mr850 humidifier, which was powered on and humming v ventilator which was in standby mode (no air flow).The setup also included an impedance valve on one of the chamber ports.A lot check revealed no other similar complaints for lot number 130519.Conclusion: it is most likely that the deformation of the chamber dome is due to the restricted flow caused by the humming v ventilator being in standby mode.The user instructions were not followed as the humidifier should not be turned on when there is no gas flow.No air flow through the chamber results in excessive heat inside the chamber which may cause the ports to deform.The weight of the impedance valve will also have caused at least one of the ports to deform.As part of our manufacturing process, every mr290 chamber undergoes pressure testing for potential leaks and those that fail are rejected.In addition, a visual inspection is performed after the port caps are assembled on to the chamber on the production line.Any product which fails this visual inspection is rejected.The user instructions that accompany the mr850 humidifier state: - "if gas flow is interrupted turn the humidifier off.".
|