Autopulse platform (s/n (b)(4)) was returned to zoll for evaluation.Visual inspection of the returned platform was performed and found no anomalies.Functional testing of the platform was performed with no issues.A review of the platform archives was performed and found no anomalies.Please note, during this review, it was found that the platform was last used on 2/18/2014 and not on the reported event date of (b)(6) 2014.The die cast channel was replaced, remedying the customer's reported complaint that the "lifeband will not stay in place".The system underwent and meets all test criteria.In conclusion, the customer's reported complaint of the 'tabs that hold the lifeband being worn" cannot be confirmed, as the lifeband was not returned with the platform for investigation.However, the customer reported complaint that the "lifeband would not stay in place" was confirmed through evaluation of the returned platform.The die cast channel was replaced, after which the platform functioned as intended.
|