• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTUITIVE SURGICAL,INC. PROGRASP FORCEPS INSTRUMENT; ENDOSCOPIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTUITIVE SURGICAL,INC. PROGRASP FORCEPS INSTRUMENT; ENDOSCOPIC INSTRUMENT Back to Search Results
Model Number 420093-08
Device Problem Malposition of Device (2616)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/27/2014
Event Type  malfunction  
Event Description
It was reported that during a da vinci cystectomy surgical procedure, the prograsp forceps instrument did not respond properly to the master control movements.The instrument felt hard to the endowrist movements.There was no report of fragments falling into a patient.There was no patient harm, adverse outcome or injury reported.
 
Manufacturer Narrative
The instrument was returned and evaluated.Failure analysis found the instrument's grip cable to be derailed.The instrument could not be placed on an in-house da vinci system due to the instrument being expired.The instrument was tested manually and one of the grip cable did not move smoothly.The housing was opened to find that one of the grip cable was derailed from the clamping pulley.Motion might not be precise or smooth due to cable derailment.Failure analysis investigation also found the instrument's pitch up cable to be frayed at the distal clevis hub.The other cables at the wrist of the instrument were undamaged.The instrument's main tube had scratch marks/abrasions.The distal end of the main tube had various scratches with light material missing.No other damage found.The customer reported complaint does not in itself constitute a mdr reportable event; however, the frayed cable and various scratch marks on the main tube with light material missing found during failure analysis if to recur could cause or contribute to an adverse event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRASP FORCEPS INSTRUMENT
Type of Device
ENDOSCOPIC INSTRUMENT
Manufacturer (Section D)
INTUITIVE SURGICAL,INC.
sunnyvale CA
Manufacturer (Section G)
INTUITIVE SURGICAL, INC.,
950 kifer rd
sunnyvale CA 9408 6
Manufacturer Contact
tabitha reed
950 kifer rd
sunnyvale, CA 94086
MDR Report Key3760009
MDR Text Key4553649
Report Number2955842-2014-02376
Device Sequence Number1
Product Code NAY
Combination Product (y/n)N
PMA/PMN Number
K050369
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Other
Type of Report Initial
Report Date 04/02/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/18/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number420093-08
Device Lot NumberM11121101 129
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/02/2014
Is the Reporter a Health Professional? No
Event Location Hospital
Date Manufacturer Received04/02/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/01/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age77 YR
-
-