Catalog Number ARC2220P
Device Problems Fluid/Blood Leak (1250); Device Inoperable (1663)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/28/2014
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that a three way stopcock leaked.This malfunction occurred during infusion on a patient.There was no report of patient injury, medical intervention, or adverse event in association with this event.No additional information is available.
 
Manufacturer Narrative
(b)(4).Initial reporter: international phone number: (b)(6).A review of all batch record documents was performed with no issues noted during the manufacturing process.There were no deviations from standard procedure and no exceptions related to the reported condition were noted.The sample was returned for evaluation.A visual inspection and leak testing were performed.During the leak test it was identified that there was a leak in the assembly between the tube and three-way clamp.The cause of the leak was a solvent dispenser failure during assembly.A capa has been opened to address this issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCESS
Type of Device
STOPCOCK, I.V. SET
Manufacturer (Section D)
BAXTER HEALTHCARE - CALI
apartado aereo 2446
calle #36
cali
CO 
Manufacturer (Section G)
BAXTER HEALTHCARE - CALI
apartado aereo 2446
calle #36
cali
CO  
Manufacturer Contact
christina arnt
25212 w. illinois route 120
round lake, IL 60073
2242703198
MDR Report Key3761470
MDR Text Key11766009
Report Number1416980-2014-12787
Device Sequence Number1
Product Code FMG
Combination Product (y/n)N
Reporter Country CodeCO
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/28/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/21/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date10/31/2016
Device Catalogue NumberARC2220P
Device Lot NumberSX13JB8
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/01/2014
Is the Reporter a Health Professional? No
Date Manufacturer Received03/28/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/03/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-