Medtronic received information that during a veno-arterial extracroporeal membrane oxygenation (ecmo) procedure, using this arterial bio-medicus (8f) cannula, the cannula appeared to be leaking.The customer reported that when the cannula was fixed with a silk tie per normal procedure the tie appeared to cut into the plastic of the lumen of the cannula.The cut was between the metal reinforced rings and reduced the lumen.There was no interruption to blood flow, but a small bead of blood appeared on the tie, which suggested that the integrity of the wall of the cannula had been breached.A decision was made to change the cannula.The patient was taken off ecmo for less than two minutes and suffered no adverse effects.The cannula was replaced and ecmo recommenced.On securing the second arterial cannula a band was placed around the cannula, and the suture was then tied onto the band so that any tightening or pressure was not applied directly onto the body of the cannula.It was reported that the surgeon is not sure of the cause of the leak but thinks the tie may have been inadvertently tied too tight.The facility is reviewing their procedure regarding securing arterial cannula.The device was discarded and will not be returned for analysis.
|