Catalog Number QD8-4B
Device Problems Metal Shedding Debris (1804); Noise, Audible (3273)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/01/2014
Event Type  Injury  
Event Description
This is report 3 of 4 of the same event.It was reported that during an unspecified foot surgery, it was observed that the motor device, attachment device and cutter device ¿were making funny noises and metal shavings were falling into the surgical site¿ when in use together.The reporter stated that the motor device was switched with a spare motor device.The same attachment device and cutter device were utilized with the spare motor device.According to the report, it was again observed that the devices were making funny noises and metal shavings were falling into the surgical site¿.As a result, there was a five minute delay to the surgical procedure.There was patient involvement.There were no reports of injuries or prolonged hospitalization.It was unknown to the reporter if radiology imaging was utilized.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
The actual device was returned for evaluation.Reliability engineering evaluated the device and the reported condition was not duplicated or confirmed.Therefore, an assignable root cause was not determined.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4MM FLUTED BALL
Type of Device
MOTOR, DRILL, ELECTRIC - CUTTING BURR
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
rodney crain
4500 riverside drive
palm beach gardens, FL 33410
5616271080
MDR Report Key3771077
MDR Text Key20714858
Report Number1045834-2014-11838
Device Sequence Number1
Product Code HBC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK011444
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/01/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/24/2014
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberQD8-4B
Device Lot NumberF023060230
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/25/2014
Is the Reporter a Health Professional? No
Date Manufacturer Received08/01/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/20/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
ATTACHMENT DEVICE; MOTOR DEVICE; MOTOR DEVICE
Patient Outcome(s) Required Intervention;
-
-