Conclusion: root cause identified in capa was found to be premature failure/saturated oil mist filter or vacuum pump oil.Asp investigation summary: the investigation included a review of the device history record, service history, trending of the product malfunction code, failure mode and effects analysis, health hazard evaluation, system hazard and user misuse analysis, and capa.The dhr (device history record) was reviewed and no issues relating to the failure mode were noted.The involved unit met manufacturer specifications at the time of release.The service history for this unit for the past 6 months (10/05/2013 to 04/03/2014) did not reveal a significant trend for this same issue.The trend of the product malfunction code odor/smells was completed from september 2012 through august 2013 and revealed a significant trend which was addressed through capa.The trending for problem code respiratory reaction (may 2013 through april 2014) did not reveal a significant trend.The trending for problem code human reaction (may 2013 through april 2014) revealed the highest risk is considered broadly acceptable.The fmea (failure mode and effects analysis) revealed the risk priority number for this failure mode is (b)(4) which is greater than the acceptable limit.A capa was opened to address this issue.The hhe (health hazard evaluation) was reviewed for the risk of odor and smell exposure.The severity and occurrence for the general population were limited (transient, minor impairment, no medical treatment required) and the product problem has been known to result in the identified harm, but only occasionally and/or under unusual circumstances.The shuma (system hazard and user misuse analysis) determined the risk is as low as reasonably practicable for exposure to odor or odorants.The capa (corrective and preventative action) identified the root cause for the odor/smells issue as: premature failure of the used and saturated sterrad® 200 oil mist filter caused oil vapor emissions that exacerbated the odor/smell complaints reported for the sterrad® 200 system.The use of a less oxidatively stable vacuum pump oil caused the odor/smells for the sterrad® 200 system.The catalytic converter was visually inspected and was found to have a broken top part.The reason for return of the catalytic converter was confirmed.The oil mist filter was returned and tested.The oil mist filter produced a significant amount of odor during pump down.The reason for return of the oil mist filter was confirmed.The vacuum pump oil was not returned.Asp will continue to track and trend this issue.
|