During an idiopathic ventricular tachycardia procedure, it was reported that the navi-star¿ thermo-cool¿ electrophysiology catheter stopped deflecting when attempting to deflect it.The catheter was replaced with a same like product.The procedure was continued and completed successfully with no patient consequences.On (b)(6), our failure analysis lab found char on the proximal end of the tip dome from the product returned, measuring more than 1 mm in length.Based on this finding, it was determined for this complaint to be reportable due to the potential risk for the patient.Awareness date changed from (b)(6).
|
(b)(4) during an idiopathic ventricular tachycardia procedure, it was reported that the navi-star thermo-cool electrophysiology catheter stopped deflecting when attempting to deflect it.The catheter was replaced with a same like product.The procedure was continued and completed successfully with no patient consequences.On april 15th, our failure analysis lab found char on the proximal end of the tip dome from the product returned, measuring more than 1 mm in length.Upon receiving, the catheter was visually inspected and char was found on the tip electrode.This condition was not reported by the customer.Per the complaint, the catheter was tested for deflection and failed.Afterwards, the catheter was dissected and the puller wire was found broken inside the handle, causing the deflection issue.Due to the char found, the catheter was tested for electrical performance, temperature response and stockert compatibility and found within specifications.In addition, an irrigation test was performed and catheter passed.The device history record (dhr) was reviewed and no anomalies were found related to this complaint.In addition, the dhr review verifies that the device was manufactured in accordance with documented specification and procedures the customer complaint was confirmed.For the char found, it remains unknown the origin of it.
|