STRYKER ORTHOPAEDICS-MAHWAH UNIVERSAL IMPACTOR/POSTIONER; INSTRUMENT Back to Search Results
Catalog Number 2101-0200
Device Problem Mechanical Problem (1384)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/10/2014
Event Type  malfunction  
Event Description
It was reported that the impactor was found to have a cross threaded end after returning to the warehouse.
 
Manufacturer Narrative
When completed, the investigation results will be submitted in a supplemental report.
 
Manufacturer Narrative
A review of the device history records indicates that the reported devices were manufactured and accepted into final stock with no reported discrepancies.The complaint history review indicated that there were no similar events for the reported lot.The threads on the device are damaged, confirming the reported event.The damaged device was discovered during inspection; there was no surgical procedure associated with the reported event.This event meets the definition of preventive maintenance; no further investigation is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.Product surveillance will continue to monitor for trends.
 
Event Description
It was reported that the impactor was found to have a cross threaded end after returning to the warehouse.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIVERSAL IMPACTOR/POSTIONER
Type of Device
INSTRUMENT
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
keyla navedo
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key3781174
MDR Text Key4445001
Report Number0002249697-2014-01679
Device Sequence Number1
Product Code MAY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123604
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/14/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/30/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number2101-0200
Device Lot NumberSMM7E00
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/30/2014
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/28/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/03/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-