BOSTON SCIENTIFIC - COSTA RICA (HEREDIA) BLAZER PRIME? XP; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER
|
Back to Search Results |
|
Model Number M004EPTP4500THN40 |
Device Problem
No Device Output (1435)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/16/2012 |
Event Type
malfunction
|
Manufacturer Narrative
|
Age at time of event: 18 years or older.Device evaluated by manufacturer: examination of the returned complaint device revealed that the catheter was broken in the torque hole area (13.6cm from the tip).Ablation testing was performed and a d-06 sensor failure was displayed.Electrical testing was performed thermistor resistance revealed an open circuit, this was due to the broken catheter, since the cables were broken at the torque hole section.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical/procedural factors.(b)(4).
|
|
Event Description
|
Reportable based on analysis completed on (b)(4) 2014.It was reported that during an ablation treatment procedure, no signal was displayed.This blazer prime xp catheter was selected; however, the catheter was connected to the cable system and no signal was displayed.The procedure was completed with another of the same device.No patient complications were reported and the patient's status is stable.However, device analysis revealed that the catheter was broken in the torque hole area.
|
|
Search Alerts/Recalls
|
|
|