Analysis of the instrument and instrument data indicate that the cause for the biased high hb1c results is user error.The account failed to input the correct lot specific scaler for one of the five scalers for the new lot of hb1c reagent.The falsely elevated results were caused by using the incorrect scaler and the issue was resolved by inputting the correct scaler d value, recalibrating, and verifying by qc.The hb1c method uses an additional parameter called scaler values.These values are polynomial equation factors that have been determined for hb1c in order to provide the best correlation to the hba1c reference methodology.This feature was communicated to customers with the launch of hb1c ((b)(4)) in the hb1c kit supplement.The instrument is performing within specifications.No further evaluation of the device is required.
|