Catalog Number 0702001000 |
Device Problem
Electrical /Electronic Property Problem (1198)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 04/09/2014 |
Event Type
malfunction
|
Event Description
|
It was reported that the neptune 2 rover ultra (120v) had bare wires showing through the insulation of the power cord during repair by a stryker field service technician.There was no patient involvement and no adverse consequences associated with the device.
|
|
Event Description
|
It was reported that the neptune 2 rover ultra (120v) had bare wires showing through the insulation of the power cord during repair by a stryker field service technician.There was no patient involvement and no adverse consequences associated with the device.
|
|
Manufacturer Narrative
|
The field service technician was able to confirm the reported event.He found that the power cord had exposed wires.The root cause was determined to be due to the power cord.The power cord¿s insulation can break down exposing the bare copper wires due to being exposed to rough surfaces, sharp edges, or corners.The device was repaired at the customer site.
|
|