Catalog Number MX0100
Device Problem Other (for use when an appropriate device code cannot be identified) (2203)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/24/2014
Event Type  malfunction  
Event Description
It was reported that a patient underwent a laparoscopic supracervical hysterectomy on (b)(6) 2014.During the procedure, the blade would not cut through the tissue.The device was making unusual noises from start and it would not cut the fibroid.Another like device was used to complete the procedure with no adverse patient consequences.
 
Manufacturer Narrative
(b)(4).The actual device involved in this event was returned for evaluation.The device was visually and functionally evaluated.Upon evaluation,the blade failed to rotate; it is likely that the accumulation of blood, body fluids, and tissue prevented the device from continuing to operate as intended.In addition, a review of the batch manufacturing records was conducted and the batch met all finished goods release criteria.
 
Manufacturer Narrative
(b)(4).Conclusion: the product upon which this medwatch is based has been received, however, the product evaluation is not yet complete.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GYNECARE MORCELLEX TISSUE MORCELLATOR
Type of Device
LAPAROSCOPE, GYNECOLOGIC (AND ACCESSORIES)
Manufacturer (Section D)
ETHICON INC.
p.o. box 151, route 22 west
somerville NJ 08876 015
Manufacturer (Section G)
MEDTECH GROUP INC
6 century ln
south plainfield NJ 07080
Manufacturer Contact
guillermo villa
route 22 west po box 151
somerville, NJ 08876
9082180707
MDR Report Key3792916
MDR Text Key4369040
Report Number2210968-2014-05770
Device Sequence Number1
Product Code HET
Combination Product (y/n)N
PMA/PMN Number
K100280
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/26/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/06/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2015
Device Catalogue NumberMX0100
Device Lot NumberMT218560
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer04/18/2014
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received05/07/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/23/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-