Device is an instrument and is not implanted/explanted.The investigation could not be completed; no conclusion could be drawn, as no product was received.Device was used for treatment, not diagnosis if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
Device was used for treatment, not diagnosis.Service history of the past six months from the awareness date was reviewed.No service history review can be performed.The item has not previously been in for service.There is no information relevant to the current complained issue.The manufacture date of this item is 23-feb-2011.The source of the manufacture date is the device history record review.Power tool evaluation was completed for this device.The customer's complaint that the (009550) battery casing does not hold charge was confirmed.The unit was tested and was found to have corrosion on the contacts.This is most likely due to immersion during cleaning.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|