• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FORNISEE SYSTEM UTERINE MANIPULATOR; CULDOSCOPE AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FORNISEE SYSTEM UTERINE MANIPULATOR; CULDOSCOPE AND ACCESSORIES Back to Search Results
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  No Answer Provided  
Event Description
(b)(4).
 
Manufacturer Narrative
On (b)(4) 2014, we were informed that a uterine perforation occurred during a total laparoscopic hysterectomy involving the fornisee system uterine manipulation product.The surgeon had experience with the fornisee technology with three prior robotic hysterectomies.In those initial cases, the physician assistant at the operating room tableside manually controlled this hand-operated device, while the surgeon performed the dissection away from the pt at the robotic console.This event involved the surgeon's first laparoscopic, non-robotic, surgery with this product; here the surgeon himself held and controlled the fs device with one hand and performed the dissection with the other.We were told that this pt has survived but experienced significant bleeding and required intervention.The fornisee product includes a metal reusable sound, called the fs sound, the distal end of which is passed preoperatively through the cervical os, cervical canal and into the uterine fundus.The fs sound has a distal end that is rotated 90 degrees after insertion into the uterine cavity.The plastic handle, called the fs device, is slid over the fs sound and through the pt's vaginal canal.The fs device has a cervical cup feature to engage the cervix and provide illumination at the area of colpotomy dissection.Reportedly, after initiating laparoscopy, some anterior wall and omental adhesions were taken down.Some dissection may have occurred near the left round ligament but none at the right round ligament.No broad ligament or tubo-ovarian ligament dissection for uterine mobilization was observed on either side.The surgeon completed the bladder flap dissection.Subsequently, the fs device's distal fully rotated end was observed outside of the uterine body.According to their medwatch report, bleeding was noted and attributed to an unobserved assessment that this uterine perforation event also "perforated a venous structure." immediate conversion to laparotomy occurred.A general surgeon was brought in.They reported, "at this point, good hemostasis was achieved and it was discussed/decided to proceed with the hysterectomy.Upon the closure of the vaginal cuff, the appearance of venous blood was noted from the previous vessel dissection site and pressure was again applied." while we have had limited communication with this hosp, we were told that the pt had continued to improve.The hosp has not released the fs sound for our eval.Based on our understanding of this event, we do not believe this represents a device malfunction.(b)(4).We believe this unfortunate event is not due to a technology design or mfg problem, but rather a known risk related to the use of uterine sounds and manipulators in gynecologic surgery.Fortunately, the pt appears to be healing well.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FORNISEE SYSTEM UTERINE MANIPULATOR
Type of Device
CULDOSCOPE AND ACCESSORIES
Manufacturer Contact
kevin bentley
7796 victor-mendon rd.
victor, NY 14564
5858696669
MDR Report Key3807873
MDR Text Key4387944
Report Number1320468-2014-00012
Device Sequence Number1
Product Code HEW
Combination Product (y/n)N
PMA/PMN Number
K111014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/15/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Other
Date Manufacturer Received03/24/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
-
-