Model Number M00542251
Device Problems Misfire (2532); Appropriate Term/Code Not Available (3191)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/17/2014
Event Type  malfunction  
Event Description
Note: this report pertains to one of four devices used during the same procedure.Refer to manufacturer report # 3005099803-2014-01943, manufacturer report # 3005099803-2014-01944, manufacturer report # 3005099803-2014-01945 and manufacturer report # 3005099803-2014-01955 for the associated device information.It was reported to boston scientific corporation that four speedband superview super7 devices were used during an upper endoscopy procedure performed on (b)(6) 2014.Reportedly the varices were located at the esophagogastric junction.According to the complainant, during the procedure, the first speedband superview super7 was able to release the bands but they missed the target area.The physician then proceeded to use the second and third speedband superview super7 devices but the same problem occurred.When the fourth speedband device was attempted to be used, the tripwire broke.The procedure was completed with a different device.There were no patient complications reported as a result of this event.The patients' condition at the conclusion of the procedure was reported to be fine.
 
Manufacturer Narrative
Reported event of bands misfired.The complainant was unable to provide the suspect device lot number; therefore, the lot expiration and device manufacture dates are unknown.However, the complainant reported that the device was not expired.The complainant indicated that the device was disposed and will not be returned for evaluation; therefore, a failure analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPEEDBAND SUPERVIEW SUPER 7?
Type of Device
LIGATOR, HEMORRHOIDAL
Manufacturer (Section D)
BOSTON SCIENTIFIC - MARLBOROUGH
100 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC - SPENCER
780 brookside drive
spencer IN 47460
Manufacturer Contact
ingrid matte
100 boston scientific way
marlborough, MA 01752
5086834000
MDR Report Key3808395
MDR Text Key4371911
Report Number3005099803-2014-01945
Device Sequence Number1
Product Code FHN
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K020824
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 04/25/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/13/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Nurse
Device Model NumberM00542251
Device Catalogue Number4225
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/25/2014
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-