Catalog Number 03.501.080 |
Device Problem
Sticking (1597)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Event Description
|
It was reported by the sterile processing department at the hospital that two zip-fix guns have triggers that are sticking.There were no reported events occurring during a procedure with patient involvement.This is report 1 of 2 for (b)(4).
|
|
Manufacturer Narrative
|
The device was received, the investigation could not be completed, and no conclusion could be drawn, as product is entering the complaint system.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Device was used for treatment, not diagnosis.Device is an instrument and is not implanted/explanted.No nonconformance reports were generated during production.Review of the device history record showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.The investigation could not be completed; no conclusion could be drawn, as no product was received.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Device was used for treatment, not diagnosis.Additional narrative: a product development evaluation was performed.There were no design related issues found on the returned instrument.The instrument in this complaint was not lubricated as per the instructions provided by the manufacture.The root cause is directly related to the not preformed lubrication during the clinical reprocessing of the device.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|