• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES HAGENDORF APPLICATION INSTRUMENT FOR STERNAL ZIPFIX; CERCLAGE FIXATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES HAGENDORF APPLICATION INSTRUMENT FOR STERNAL ZIPFIX; CERCLAGE FIXATION Back to Search Results
Catalog Number 03.501.080
Device Problem Sticking (1597)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Event Description
It was reported by the sterile processing department at the hospital that two zip-fix guns have triggers that are sticking.There were no reported events occurring during a procedure with patient involvement.This is report 1 of 2 for (b)(4).
 
Manufacturer Narrative
The device was received, the investigation could not be completed, and no conclusion could be drawn, as product is entering the complaint system.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Device was used for treatment, not diagnosis.Device is an instrument and is not implanted/explanted.No nonconformance reports were generated during production.Review of the device history record showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.The investigation could not be completed; no conclusion could be drawn, as no product was received.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Device was used for treatment, not diagnosis.Additional narrative: a product development evaluation was performed.There were no design related issues found on the returned instrument.The instrument in this complaint was not lubricated as per the instructions provided by the manufacture.The root cause is directly related to the not preformed lubrication during the clinical reprocessing of the device.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
APPLICATION INSTRUMENT FOR STERNAL ZIPFIX
Type of Device
CERCLAGE FIXATION
Manufacturer (Section D)
SYNTHES HAGENDORF
im bifang 6
hagendorf CH46 14
SZ  CH4614
Manufacturer (Section G)
SYNTHES HAGENDORF
im bifang 6
hagendorf CH46 14
SZ   CH4614
Manufacturer Contact
linda plews
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key3819344
MDR Text Key4396572
Report Number3003875359-2014-10122
Device Sequence Number1
Product Code JDQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK110789
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date 04/28/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/19/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number03.501.080
Device Lot Number8271971
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/07/2014
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/24/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/04/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-