• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BATTLE CREEK EQUIPMENT THERMOPHORE AUTOMATIC HEAT PACK; ELECTRIC MOIST HEAT PACK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BATTLE CREEK EQUIPMENT THERMOPHORE AUTOMATIC HEAT PACK; ELECTRIC MOIST HEAT PACK Back to Search Results
Model Number 177
Device Problems Device Emits Odor (1425); Overheating of Device (1437)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/14/2014
Event Type  malfunction  
Event Description
Customer stated that something smelled like it was burning and the pad was getting hotter than it should.
 
Manufacturer Narrative
Qc tested pad and it was heating and working properly, (136-156).No odor poly cover ok, cord ok, switch ok, pad ok.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMOPHORE AUTOMATIC HEAT PACK
Type of Device
ELECTRIC MOIST HEAT PACK
Manufacturer (Section D)
BATTLE CREEK EQUIPMENT
freemont IN
Manufacturer Contact
randy newsome
702 south reed st.
freemont, IN 46737
2699626181
MDR Report Key3823756
MDR Text Key17614711
Report Number1811605-2014-00113
Device Sequence Number1
Product Code IRT
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,User Facility
Reporter Occupation Not Applicable
Type of Report Initial
Report Date 05/01/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/02/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number177
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/25/2014
Is the Reporter a Health Professional? No
Date Manufacturer Received04/25/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/01/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-