• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORPORATION SINGLE USE 3-LUMEN SPHINCTEROTOME

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORPORATION SINGLE USE 3-LUMEN SPHINCTEROTOME Back to Search Results
Model Number KD-V411M-0720
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/15/2014
Event Type  malfunction  
Event Description
Olympus medical systems corp (omsc) was informed that during an endoscopic sphincterotomy (est) with sphincterotome, the knife could not activate second output.The doctor withdrew the device from the endoscope and then found that the knife wire broke.The doctor completed the procedure by using another device.During the investigation, omsc found the plastic coating on the cutting wire was partially missing.There was no report of patient injury regarding this report.
 
Manufacturer Narrative
The investigation also confirmed that the cutting wire was broken at the coated portion adn the broken section was melted and burned.Approximately coating was missing for 7mm from the broken point.There were no other abnormalities related to the breakage in the subject device.Also as the result of checking the manufacturing record of the same lot, nothing abnormal detected.As the results of the investigation, omsc assumes that the damage of the coating occurred due to contacting with the metal part of the forceps elevator of the endoscope.The exposed cutting wire from the damaged coating contacted or came close to the metal part of the forceps elevator while activating the output, which caused spark and a part of the cutting wire became extremely hot, resulting in breakage.The coating broke off and came off from the cutting wire because of the user handling after the cutting wire was broken.This report is being submitted as a medical device report in an abundance of caution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE 3-LUMEN SPHINCTEROTOME
Type of Device
SPHINCTEROTOME
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORPORATION
2951 ishikawa-cho
hachioji-shi, tokyo 192- 8507
JA  192-8507
Manufacturer Contact
hiroki moriyama
2951 ishikawa-cho
hachioji-shi, tokyo 192-8-507
JA   192-8507
26425177
MDR Report Key3830414
MDR Text Key4395376
Report Number8010047-2014-00217
Device Sequence Number1
Product Code FDR
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Physician
Type of Report Initial
Report Date 04/23/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/13/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberKD-V411M-0720
Device Lot NumberK4214
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer04/21/2014
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/23/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/01/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age78 YR
-
-