Catalog Number L5C4531
Device Problems Device Inoperable (1663); Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/02/2014
Event Type  malfunction  
Event Description
It was reported that the home patient (hp) had a system error 2240 (air in tubing) on the homechoice (hc) during dwell three of six.The technical service representative (tsr) had the home patient (hp) check the lines and bags, close the clamps and cycle the power.The tsr had the hp disconnect using aseptic technique and remove the cassette.There was nothing unusual found during the troubleshooting that could have caused or contributed to the reported alarm.There was no patient injury or adverse event associated with this report.No additional information is available.
 
Manufacturer Narrative
(b)(4).An alarm indicative of a potential malfunction of the disposable cassette was identified.As the cassette was not returned and the lot number was unknown, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer (Section G)
BAXTER HEALTHCARE - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer Contact
christina arnt
25212 w. illinois route 120
round lake, IL 60073
2242703198
MDR Report Key3833331
MDR Text Key4496724
Report Number1416980-2014-17035
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K923065
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 05/02/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/27/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue NumberL5C4531
Was Device Available for Evaluation? No
Date Manufacturer Received05/02/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
HOMECHOICE
Patient Age74 YR
-
-