MAUDE Adverse Event Report: BECKMAN COULTER UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM; COUNTER, DIFFERENTIAL CELL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECKMAN COULTER UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM; COUNTER, DIFFERENTIAL CELL Back to Search Results
Catalog Number 629029
Device Problems Device Emits Odor (1425); Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/30/2014
Event Type  malfunction  
Event Description
The customer reported a burning smell coming from the back of the unicel dxh 800 coulter cellular analysis system while running latron quality control (qc).The customer stated a "pneumatic supply module exceeded the operating limits" system error was also obtained.The customer powered off the instrument.There were no sparks, arcs or flames.The lab was not evacuated and the fire department was not called.No erroneous results were generated in association with the event.There was no death, injury, or affect to user or to patient treatment attributed to this event.
 
Manufacturer Narrative
A beckman coulter (bec) field service engineer (fse) was dispatched to assess the instrument.The fse found a quick disconnect 8 on the flow cell harness under the sheath tank was not tight and was leaking slightly (line provides diluent for sample line cleaning and interconnects to the sample line) onto the pneumatic services, temp control card and shorted solenoids.The fse tightened the quick disconnect and also replaced pneumatic module resource card (mrc), pneumatic sensor card, volume conductivity scatter (vcsn) temperature controller card, module id board and associated cables to resolve the issue with the solenoids damaged by the leak.The fse verified the instrument with no further issues.The instrument was returned to normal operation.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM
Type of Device
COUNTER, DIFFERENTIAL CELL
Manufacturer (Section D)
BECKMAN COULTER
250 s. kraemer blvd.
brea CA 92821
Manufacturer (Section G)
BECKMAN COULTER
11800 sw 147th avenue
miami FL 33196
Manufacturer Contact
dung nguyen
250 s. kraemer blvd.
brea, CA 92821
7149614941
MDR Report Key3840019
MDR Text Key4661670
Report Number1061932-2014-01132
Device Sequence Number1
Product Code GKZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K081930
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/01/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/30/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number629029
Other Device ID NumberSW VERSION: 3.0
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/01/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-