It was reported that during a da vinci surgical procedure, the customer noted that the pk dissecting forceps instrument had a bent tip and the tips would not meet.There was no report of fragments falling into a patient.There was no allegation of any harm, injury, or adverse outcome to a patient.
|
The instrument was returned and evaluated.Per the customer reported complaint, failure analysis investigation found that the instrument tips were bent.One grip was bent, causing side to side misalignment of grips.There was a.035 offset at the tips.The grip tips did not meet together when fully closed.The instrument passed electrical continuity testing.It was concluded that the damages to the instrument's grip tip were likely due to mishandling/misuse.An additional observation not reported by the site was that the instrument's pitch cable was frayed at the distal clevis hub.The frayed strands stuck out at the wrist.Other cables at wrist were undamaged.No other damage was found.The endowrist instruments instructions for use (ifu) specifically states: general precautions and warnings handle instruments with care.Avoid mechanical shock or stress that can cause damage to the instruments.The customer reported complaint does not itself constitute a mdr reportable event; however; the damage to the instrument's pitch cable found during failure analysis investigation could cause or contribute to an adverse event if the failure mode were to recur.
|