STRYKER INSTRUMENTS-KALAMAZOO REMB SAG SAW; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)
|
Back to Search Results |
|
Catalog Number 6400034000 |
Device Problem
Device Displays Incorrect Message (2591)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 05/09/2014 |
Event Type
malfunction
|
Event Description
|
The remb sagittal saw was returned to stryker instruments for service.During functional testing by a service technician, it was found that the device caused a bias current message to be displayed on the console, signaling a condition occurred in which the device has the potential to run without user activation.No medical intervention and no adverse consequences were reported with this event.As this event occurred during testing at the manufacturer facility, there was no patient involvement and no delay to a surgical procedure.
|
|
Manufacturer Narrative
|
The reported event, bias current error on the console, was duplicated.Upon disassembly for visual inspection the service technician observed internal motor assembly corrosion.
|
|