Catalog Number 6400034000
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Patient Involvement (2645)
Event Date 05/09/2014
Event Type  malfunction  
Event Description
The remb sagittal saw was returned to stryker instruments for service.During functional testing by a service technician, it was found that the device caused a bias current message to be displayed on the console, signaling a condition occurred in which the device has the potential to run without user activation.No medical intervention and no adverse consequences were reported with this event.As this event occurred during testing at the manufacturer facility, there was no patient involvement and no delay to a surgical procedure.
 
Manufacturer Narrative
The reported event, bias current error on the console, was duplicated.Upon disassembly for visual inspection the service technician observed internal motor assembly corrosion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REMB SAG SAW
Type of Device
DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)
Manufacturer (Section D)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer (Section G)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer Contact
casey metzger
4100 east milham avenue
kalamazoo, MI 49001
2693237700
MDR Report Key3845045
MDR Text Key4687539
Report Number0001811755-2014-02025
Device Sequence Number1
Product Code HBE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K032303
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 05/09/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/03/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number6400034000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/01/2014
Date Manufacturer Received05/09/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/27/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse