Device Problem Loss of Power (1475)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/12/2014
Event Type  No Answer Provided  
Event Description
The user facility reported that their 3085 surgical table was losing power during a patient procedure.User facility personnel were able to use the auxiliary switches to complete the procedure successfully.No injuries or procedural delays/cancellations were reported.
 
Manufacturer Narrative
A steris service technician inspected the table and found the hand control was indicating low battery power.He noted that one of the leads on the control batteries had some corrosion which likely caused the reported loss or intermittent power issues.The technician replaced the control batteries and as a precautionary measure replaced the power supply.The technician tested the table and returned it to service; no additional issues have been reported.The table is not under steris service contract and is serviced and maintained by the user facility.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3085 SURGICAL TABLE
Type of Device
SURGICAL TABLE
Manufacturer (Section D)
STERIS CORPORATION - MONTGOMERY
2720 gunter park drive east
montgomery AL 36109
Manufacturer (Section G)
STERIS CORPORATION - MONTGOMERY
2720 gunter park drive east
montgomery AL 36109
Manufacturer Contact
kathryn cadorette
5960 heisley road
mentor, OH 44060
4403927231
MDR Report Key3848961
MDR Text Key4687124
Report Number1043572-2014-00053
Device Sequence Number1
Product Code FQO
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 06/04/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/04/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/12/2014
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-