Device Problem
Loss of Power (1475)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 05/12/2014 |
Event Type
No Answer Provided
|
Event Description
|
The user facility reported that their 3085 surgical table was losing power during a patient procedure.User facility personnel were able to use the auxiliary switches to complete the procedure successfully.No injuries or procedural delays/cancellations were reported.
|
|
Manufacturer Narrative
|
A steris service technician inspected the table and found the hand control was indicating low battery power.He noted that one of the leads on the control batteries had some corrosion which likely caused the reported loss or intermittent power issues.The technician replaced the control batteries and as a precautionary measure replaced the power supply.The technician tested the table and returned it to service; no additional issues have been reported.The table is not under steris service contract and is serviced and maintained by the user facility.
|
|