• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORP. TERUMO TCM II COOLING AND HEATING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORP. TERUMO TCM II COOLING AND HEATING SYSTEM Back to Search Results
Model Number 4415
Device Problems Partial Blockage (1065); Device Operates Differently Than Expected (2913)
Patient Problem Not Applicable (3189)
Event Date 01/01/2013
Event Type  malfunction  
Event Description
It was reported that during the use of the device for a non-clinical activity, an ice block froze over all of the water in the tank of the cooler heater unit.It was discovered during daily chlorination testing.The unit was allowed to thaw.Once thawed, the unit was rinsed and flushed.The unit was put back into service and has not malfunctioned since per the second perfusionist (ccp).There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TERUMO TCM II COOLING AND HEATING SYSTEM
Type of Device
TCM II COOLING AND HEATING SYSTEM
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORP.
ann arbor MI 48103
Manufacturer Contact
jan winder
6200 jackson rd.
ann arbor, MI 48103
7346634145
MDR Report Key3851104
MDR Text Key4413541
Report Number1828100-2014-00092
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K883603
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/21/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/12/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number4415
Device Catalogue Number4415
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/21/2014
Was Device Evaluated by Manufacturer? No
Date Device Manufactured08/01/2003
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-