• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MED REL MEDTRONIC PUERTO RICO ADAPTA SR; PULSE GENERATOR, PERMANENT, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MED REL MEDTRONIC PUERTO RICO ADAPTA SR; PULSE GENERATOR, PERMANENT, IMPLANTABLE Back to Search Results
Model Number ADSR01
Device Problem Material Integrity Problem (2978)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Event Description
The device was returned to the manufacturer, analyzed, and tested out of specification.Additional information obtained indicated the device was explanted post-mortem and the cause of death was not related to the out of specification finding.
 
Manufacturer Narrative
Product event summary: the device was returned and analyzed.Analysis revealed there was no output and no telemetry due to a defective crystal oscillator.The information submitted reflects all relevant data received.If additional relevant information is received, a supplemental report will be submitted.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADAPTA SR
Type of Device
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
ceiba norte industrial park, r
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM DISEASE MGMT
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
nashoane fulwood-kelley
8200 coral sea st ne
mounds view, MN 55112
7635260583
MDR Report Key3858187
MDR Text Key19968643
Report Number3004209178-2014-10380
Device Sequence Number1
Product Code NVZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P980035
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 04/03/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/09/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/14/2014
Device Model NumberADSR01
Device Catalogue NumberADSR01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/26/2013
Is the Reporter a Health Professional? No
Date Manufacturer Received04/03/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/15/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
4092-52 LEAD
Patient Age00083 YR
-
-