Model Number 6215 |
Device Problems
Mechanical Problem (1384); Activation, Positioning or Separation Problem (2906)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 03/07/2014 |
Event Type
malfunction
|
Event Description
|
It was reported that at the implant procedure the physician had difficulty advancing the lead due to tortuous anatomy.Upon examination, the physician thought the lead was distorted or "off center." a new lead was implanted.It was also reported that the valve at the proximal end of the catheter was deformed.The physician had difficulty getting the guidewire into the catheter and advancing it down to the tip.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
The information submitted reflects all relevant data received.If additional relevant information is received, a supplemental report will be submitted.Product event summary: the catheter was returned and analyzed.The luer loc is flattened.Guidewire passed thru catheter and out the tip end.(b)(4).
|
|