• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC, INC. ATTAIN VENOGRAM BALLOON CATHETER; CATHETER, FLOW DIRECTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC. ATTAIN VENOGRAM BALLOON CATHETER; CATHETER, FLOW DIRECTED Back to Search Results
Model Number 6215
Device Problems Mechanical Problem (1384); Activation, Positioning or Separation Problem (2906)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/07/2014
Event Type  malfunction  
Event Description
It was reported that at the implant procedure the physician had difficulty advancing the lead due to tortuous anatomy.Upon examination, the physician thought the lead was distorted or "off center." a new lead was implanted.It was also reported that the valve at the proximal end of the catheter was deformed.The physician had difficulty getting the guidewire into the catheter and advancing it down to the tip.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
The information submitted reflects all relevant data received.If additional relevant information is received, a supplemental report will be submitted.Product event summary: the catheter was returned and analyzed.The luer loc is flattened.Guidewire passed thru catheter and out the tip end.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTAIN VENOGRAM BALLOON CATHETER
Type of Device
CATHETER, FLOW DIRECTED
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC CARDIAC RHYTHM DISEASE MGMT
8200 coral sea st ne
mounds view MN 55112
Manufacturer Contact
nashoane fulwood-kelley
8200 coral sea st ne
mounds view, MN 55112
7635260583
MDR Report Key3860612
MDR Text Key4687296
Report Number2182208-2014-01468
Device Sequence Number1
Product Code DYG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K012225
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/07/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/10/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/04/2015
Device Model Number6215
Device Catalogue Number6215
Device Lot Number61331714
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/17/2014
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/07/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/27/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age00051 YR
-
-