• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER MEDICAL-KALAMAZOO CAPITAL TEMP PUMP PROFESSIONAL; PACK, HOT OR COLD, WATER CIRCULATING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER MEDICAL-KALAMAZOO CAPITAL TEMP PUMP PROFESSIONAL; PACK, HOT OR COLD, WATER CIRCULATING Back to Search Results
Catalog Number TP700
Device Problem Grounding Malfunction (1271)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/12/2014
Event Type  malfunction  
Event Description
It was reported that the unit failed the ground resistance portion of the functional testing.The unit was replaced.No patient was affected and no adverse consequence or clinically relevant delay in treatment was reported.
 
Event Description
It was reported that the unit failed the ground resistance portion of the functional testing.The unit was replaced.No patient was affected and no adverse consequence or clinically relevant delay in treatment was reported.
 
Manufacturer Narrative
The customer was sent a replacement unit.
 
Manufacturer Narrative
The issue was resolved for the customer by replacing the motor and mounting plate and returning the unit to the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPITAL TEMP PUMP PROFESSIONAL
Type of Device
PACK, HOT OR COLD, WATER CIRCULATING
Manufacturer (Section D)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer (Section G)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer Contact
brian thompson
3800 east centre avenue
portage, MI 49002
2693292100
MDR Report Key3864170
MDR Text Key4510169
Report Number0001831750-2014-03054
Device Sequence Number1
Product Code ILO
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/15/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/10/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberTP700
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received05/15/2014
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-