The manufacturer has investigated this case and the log files appear to indicate that an actual mistreatment occurred.The fact that the lateral position is identical to the previous scan, when taken in conjunction with the theraview records, would appear to indicate that the xvi table lateral position was frozen ie the lateral position was not being updated as the table was moved.If we assume that the theraview lateral position of 0.14 was correct, then treatment occurred approx 6.5cm away from the correct position.Field safety corrective action ((b)(4)) was released (b)(4) 2014 for distribution to customers.
|