BECKMAN COULTER UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM; COUNTER, DIFFERENTIAL CELL Back to Search Results
Catalog Number 629029
Device Problems Leak/Splash (1354); Device Contamination with Body Fluid (2317)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/28/2014
Event Type  malfunction  
Event Description
The customer reported blood on top of the sample tubes, and a possible leak, when using the unicel dxh 800 coulter cellular analysis system.A beckman coulter field service engineer (fse) was sent to the customer's facility to evaluate the instrument.The fse identified a leak of approximately 0.5ml from the instrument.The leak was not contained within the instrument.The operator was wearing personal protective equipment.There were no reports of biohazard exposure to mucous membranes or cuts.There were no erroneous test results associated with this event.There was no death, injury or affect to user or patient treatment.
 
Manufacturer Narrative
On (b)(4) 2014, the field service engineer (fse) evaluated the instrument and identified a leak caused by insufficient vacuum supplied to the probe wipe.Solenoid 341, (vl341) was replaced to resolve the issue.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM
Type of Device
COUNTER, DIFFERENTIAL CELL
Manufacturer (Section D)
BECKMAN COULTER
250 s. kraemer blvd.
brea CA 92821
Manufacturer (Section G)
BECKMAN COULTER
11800 sw 147th avenue
miami FL 33196
Manufacturer Contact
dung nguyen
250 s kraemer blvd
brea, CA 92821
7149614941
MDR Report Key3885967
MDR Text Key4545843
Report Number1061932-2014-01407
Device Sequence Number1
Product Code GKZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K081930
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/28/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/20/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number629029
Other Device ID Number2.0.2.1
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/28/2014
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/01/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-