Model Number 57062/9207010
Device Problems Crack (1135); Sparking (2595)
Patient Problem No Information (3190)
Event Date 04/14/2014
Event Type  malfunction  
Event Description
Customer reported that the transformer sparked when pulled out from the wall and the housing was cracked and she could see exposed wires which is a safety risk.
 
Manufacturer Narrative
A replacement power supply was sent to the customer.In follow up with the customer she stated that parts of the plastic housing cracked and fell away from the unit and she could see exposed wires which is a safety risk.She confirmed that there was no injury from the event.No smoke or fire, but there was a spark when she pulled the unit from the wall at the prongs.This issue with the damaged transformer housing for the pump in style device was addressed in investigation (b)(4).The investigation found that the transformers are being damaged during shipment from the manufacturer to medela.This damage is causing the plastic housing to fall prematurely when subjected to normal use and foreseeable misuse.The packaging used by the manufacturer to ship the transformers to medela is not robust enough to handle all of the potential shipping, handling, and abuse conditions that could arise from logistics of the consolidation process.As a result of the investigation, the shipping and consolidation process has been modified to reduce the handling and potential for double stacking of the skids.The shipping packaging strength has also been increased to further protect the transformers during shipping.This corrective action applies to transformers with a date code greater than or equal to 3912.Complaint data trending will continue to be monitored by medela quality management for investigation/capa consideration.Possibility resolutions provided to customer: replacement product or online ordering information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PUMP IN STYLE ADVANCED BREAST PUMP BACKPACK
Manufacturer (Section D)
MEDELA, INC.
mchenry IL 60050
Manufacturer Contact
don alexander
1101 corporate dr.
mchenry, IL 60050
8004355316
MDR Report Key3888121
MDR Text Key20104141
Report Number1419937-2014-00374
Device Sequence Number1
Product Code HGX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K031614
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Not Applicable
Type of Report Initial
Report Date 04/14/2014
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/21/2014
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number57062/9207010
Device Catalogue Number57062/9207010
Device Lot NumberREV L - 1811
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received04/14/2014
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-