Dhr records were verified and no discrepancy was recorded on these items.Raw materials inspection records with no discrepancy.(b)(4).Concentricity of 1mm hole was found within spec.No other defects were detected.According to the attached image the fracture was caused due to wrong bending conditions and extensive external load.This device intended to be used on soft tissue as stated in the ifu.In the report it is stated that device was used on labrum and that the surgeon likely struck bone.This device is not designed to be in use on bones.See the following in the ifu 505st - paragraph indication and precautions.This product use is passing suture tissue during minimally invasive surgery.Do not apply excessive axial or bending force to the slider needle during use, as excessive force may limit the function of the device or cause the device to bend, deform or break.
|