(b)(4).Method: the complaint rt134 breathing circuit was visually inspected, pressure tested and immersed in a water bath to test for leaks.Results: pressure testing revealed that the circuit was out of specification.The water bath test revealed a leak between the water trap bowl and the lid.A lot check was unable to be performed as no lot number was provided by the hospital.Conclusion: the water trap of the breathing circuit consists of two parts, a lid and a bowl which can be disconnected during use for draining condensate.The water trap leak was located at the seal between the water trap bowl and the water trap lid.All circuits are pressure tested for leaks during production and those that fail are rejected.This suggests any leak must have developed post production during transport, storage or use, possibly by distortion of the water trap when the bowl was connected.The customer had commented that the complaint circuits passed the ventilator leak test which suggests that the water traps became loose during setup of the devices.The user instructions that accompany the rt134 adult breathing circuit state the following: - perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient.- set appropriate ventilator alarms.
|